Migration and sac expansion as modes of midterm therapeutic failure after endovascular aneurysm sealing: 295 cases at a single center.

Stenson, KM; de Bruin, JL; Loftus, IM; Holt, PJE (2020) Migration and sac expansion as modes of midterm therapeutic failure after endovascular aneurysm sealing: 295 cases at a single center. J Vasc Surg, 71 (2). 457-469.e1. ISSN 1097-6809 https://doi.org/10.1016/j.jvs.2019.04.482
SGUL Authors: Stenson, Katherine Mary Loftus, Ian Holt, Peter James Edward

[img] Microsoft Word (.docx) Accepted Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.

Download (81kB)

Abstract

BACKGROUND: Endovascular aneurysm sealing (EVAS) is a disruptive technology to treat abdominal aortic aneurysm (AAA). The use of sac filling rather than endograft fixation was designed to treat aortic aneurysms in a wide range of morphologic appearances and to reduce endoleaks. There are few data reporting outcomes beyond postoperative follow-up. This study reports outcomes up to 5 years for Nellix (Endologix, Irvine, Calif) EVAS. METHODS: Data were prospectively collected for EVAS patients from the time of adoption of EVAS in 2013. All patients treated with the Nellix device are included in this study, and as such, it reports on infrarenal, ruptured, and iliac aneurysms as well as the Nellix-in-Nellix application. Juxtarenal and suprarenal aneurysms were treated using the EVAS system with parallel grafts into the visceral vessels and are included. Therapeutic failure, a composite outcome of migration, sac expansion >5 mm, type Ia and type Ib endoleak, and secondary aortic rupture, was the primary outcome along with all-cause mortality, aneurysm-related mortality, and reintervention rates. RESULTS: There were 295 EVAS cases undertaken between March 2013 and July 2018. Indications for treatment were infrarenal (n = 185), juxtarenal and suprarenal (n = 73), ruptured (n = 18), and iliac (n = 13) aneurysms. There were 15 reinterventions using the Nellix-in-Nellix application. In some cases, EVAS was used to salvage failing endovascular or open aneurysm repairs. Median follow-up was 2.42 years (interquartile range, 1.07-3.57 years). Therapeutic failure was observed in 98 of the 295 cases (33.2%) overall and exceeded 50% in some subgroups. In 71 cases (24.1%), reintervention was performed, with reasons for no reintervention being mainly physiologic. Complications leading to therapeutic failure were most commonly seen beyond 2 years of follow-up. There were 15 secondary ruptures (5.36%), and 9 EVAS devices required explantation either electively or for aortic rupture. CONCLUSIONS: EVAS with the Nellix device has not met expectations, and early encouraging results have been eroded. The incidence of therapeutic failure has been high, occurring 2 years and beyond after implantation. The Nellix system has been voluntarily recalled by Endologix, and the CE mark has subsequently been suspended. The adoption of EVAS as a disruptive technology highlights the need for cautious adoption of novel technologies and the strict governance around such arrangements.

Item Type: Article
Additional Information: © 2019. This manuscript version is made available under the CC-BY-NC-ND 4.0 license http://creativecommons.org/licenses/by-nc-nd/4.0/ Tables and figures are available via the publisher version of record.
Keywords: Abdominal aortic aneurysm, Aneurysm sealing, Endovascular, Therapeutic failure, 11 Medical And Health Sciences, Cardiovascular System & Hematology
SGUL Research Institute / Research Centre: Academic Structure > Molecular and Clinical Sciences Research Institute (MCS)
Academic Structure > Molecular and Clinical Sciences Research Institute (MCS) > Vascular & Cardiac Surgery (INCCVC)
Journal or Publication Title: J Vasc Surg
ISSN: 1097-6809
Language: eng
Dates:
DateEvent
1 February 2020Published
9 August 2019Published Online
5 April 2019Accepted
Publisher License: Creative Commons: Attribution-Noncommercial-No Derivative Works 4.0
PubMed ID: 31405762
Go to PubMed abstract
URI: https://openaccess.sgul.ac.uk/id/eprint/111115
Publisher's version: https://doi.org/10.1016/j.jvs.2019.04.482

Actions (login required)

Edit Item Edit Item