The effect of multiple analysers on the biochemical diagnosis of myocardial infarction using a contemporary troponin-I assay.

Pethick, J; Patel, P; Davies, T; Thompson, J; Nallagonda, M; Beech, A; Collinson, P; Lee, V; Gupta, P (2016) The effect of multiple analysers on the biochemical diagnosis of myocardial infarction using a contemporary troponin-I assay. Ann Clin Biochem, 53 (6). pp. 702-705. ISSN 1758-1001 https://doi.org/10.1177/0004563216637388
SGUL Authors: Collinson, Paul

[img] Microsoft Word (.docx) Accepted Version
Available under License ["licenses_description_publisher" not defined].

Download (70kB)

Abstract

Background The measurement of cardiac troponin is central for the diagnosis of myocardial infarction (MI). It is recommended that a coefficient of variation of ≤10% is achieved at the diagnostic threshold and significant change between serial measurements reported. Many modern laboratories use multiple analysers linked by automation where samples are randomly assigned to an analyser. It is therefore important to consider the combined effect of all analysers on the analytical performance of troponin measurement. Method The performance of a contemporary troponin-I (cTn-I) assay run on three analysers, linked by an automated track, was undertaken across a range of cTn-I concentrations. The data for the three analysers were aggregated to obtain the combined analytical coefficient of variation (CVA) and reference change values (RCVs). Results The CVA improved with increasing concentration and calculated RCVs ranged from 67.2% (±13 ng/L) to 32% (±160 ng/L) between cTn-I values 20 ng/L and 500 ng/L. Although there were significant differences in cTn-I measurement between analysers around the diagnostic threshold ( P < 0.05), the CVA was 13.6%. Conclusions We demonstrate that there are significant differences between the performances of analysers which can impact the biochemical criteria for the diagnosis of MI. We also show that the RCV varies according to baseline cTn-I values and that reporting a single RCV across the analytical range of cTn-I may not be appropriate.

Item Type: Article
Additional Information: Pethick, J; Patel, P; Davies, T; Thompson, J; Nallagonda, M; Beech, A; Collinson, P; Lee, V; Gupta, P (2016) The effect of multiple analysers on the biochemical diagnosis of myocardial infarction using a contemporary troponin-I assay. Ann Clin Biochem, 53 (6). pp. 702-705. Copyright © 2016. Reprinted by permission of SAGE Publications. https://doi.org/10.1177/0004563216637388
Keywords: Troponin, analytes, laboratory methods, statistics, Automation, Laboratory, Biological Assay, Biomarkers, Humans, Myocardial Infarction, Myocardium, Observer Variation, Reference Values, Troponin T, Troponin, analytes, statistics, laboratory methods, Troponin, analytes, laboratory methods, statistics, General Clinical Medicine, 1101 Medical Biochemistry And Metabolomics
SGUL Research Institute / Research Centre: Academic Structure > Molecular and Clinical Sciences Research Institute (MCS)
Academic Structure > Molecular and Clinical Sciences Research Institute (MCS) > Cardiac (INCCCA)
Journal or Publication Title: Ann Clin Biochem
ISSN: 1758-1001
Language: eng
Dates:
DateEvent
1 November 2016Published
20 April 2016Published Online
15 February 2016Accepted
Publisher License: Publisher's own licence
PubMed ID: 27099337
Web of Science ID: WOS:000385743900012
Go to PubMed abstract
URI: https://openaccess.sgul.ac.uk/id/eprint/108968
Publisher's version: https://doi.org/10.1177/0004563216637388

Actions (login required)

Edit Item Edit Item