A Prospective Study to Evaluate the Accuracy of Pulse Power Analysis to Monitor Cardiac Output in Critically Ill Patients.

Cecconi, M; Fawcett, J; Grounds, RM; Rhodes, A (2008) A Prospective Study to Evaluate the Accuracy of Pulse Power Analysis to Monitor Cardiac Output in Critically Ill Patients. BMC Anesthesiology, 8 (3). ISSN 1471-2253 https://doi.org/10.1186/1471-2253-8-3
SGUL Authors: Cecconi, Maurizio Rhodes, Andrew

[img]
Preview
["document_typename_application/pdf; charset=binary" not defined] Published Version
Download (663kB) | Preview

Abstract

BACKGROUND: Intermittent measurement of cardiac output may be performed using a lithium dilution technique (LiDCO). This can then be used to calibrate a pulse power algorithm of the arterial waveform which provides a continuous estimate of this variable. The purpose of this study was to examine the duration of accuracy of the pulse power algorithm in critically ill patients with respect to time when compared to measurements of cardiac output by an independent technique. METHODS: Pulse power analysis was performed on critically ill patients using a proprietary commercial monitor (PulseCO). All measurements were made using an in-dwelling radial artery line and according to manufacturers instructions. Intermittent measurements of cardiac output were made with LiDCO in order to validate the pulse power measurements. These were made at baseline and then following 1, 2, 4 and 8 hours. The LiDCO measurement was considered the reference for comparison in this study. The two methods of measuring cardiac output were then compared by linear regression and a Bland Altman analysis. An error rate for the limits of agreement (LOA) between the two techniques of less than 30% was defined as being acceptable for this study. RESULTS: 14 critically ill medical and surgical patients were enrolled over a three month period. At baseline patients showed a wide range of cardiac output (median 7.5 L/min, IQR 5.1 -9.0 L/min). The bias and limits of agreement between the two techniques was deemed acceptable for the first four hours of the study with percentage errors being 29%, 22%, and 285 respectively. The percentage error at eight hours following calibration increased to 36%. The ability of the PulseCo to detect changes in cardiac output was assessed with a similar analysis. The PulseCO tracked the changes in cardiac output with adequate accuracy for the first four hours with percentage errors being 20%, 24% and 25%. However at eight hours the error had increased to 43%. CONCLUSION: The agreement between lithium dilution cardiac output and the pulse power algorithm in the PulseCO monitor remains acceptable for up to four hours in critically ill patients.

Item Type: Article
Additional Information: © 2008 Cecconi et al; licence see BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Keywords: Anesthesiology, 1116 Medical Physiology
SGUL Research Institute / Research Centre: Academic Structure > Institute of Medical & Biomedical Education (IMBE)
Academic Structure > Institute of Medical & Biomedical Education (IMBE) > Centre for Clinical Education (INMECE )
Journal or Publication Title: BMC Anesthesiology
ISSN: 1471-2253
Dates:
DateEvent
February 2008Published
PubMed ID: 18282275
Go to PubMed abstract
URI: https://openaccess.sgul.ac.uk/id/eprint/107210
Publisher's version: https://doi.org/10.1186/1471-2253-8-3

Actions (login required)

Edit Item Edit Item